Full automated data processing: a case study using GPCRs
Introduction
Single particle cryo-EM is a valuable and growing technique for life science and drug discovery. Currently, obtaining state-of-the-art results from cryo-EM data analysis requires a human in the loop to analyze intermediate results and make image processing decisions. This bottleneck limits the achievable throughput of structure determination, especially in high-throughput settings such as structure-based drug design. In this work, we develop an end-to-end automation strategy for repeat-target structure determination using new tools in CryoSPARC.
We demonstrate completely hands-off processing of 21 challenging G protein-coupled receptor (GPCR) datasets. In 17 of 21 cases, automated processing meets or exceeds published resolution and map quality and, in several cases, provides significant improvement in receptor and ligand density that allows improved model building. Our results on both active and inactive state GPCRs show that our automation strategy generalizes easily to new target classes, and that complete automation of data processing is straightforward to achieve in CryoSPARC. We provide downloadable CryoSPARC Workflow files so that users can import, replicate, adapt and extend our automated workflow for their own targets, enabling cryo-EM to be applied at larger scales and to answer larger biological questions.
GPCRs: A challenging test case for automation
GPCRs are small, real-world drug targets and are generally challenging for single particle analysis. We selected 21 EMPIAR datasets of GPCR complexes in both the active and inactive state (see Table 1 in the preprint for more details).
Development of a single-particle automation strategy: Workflow v1
We detail the development of the generalized automation strategy in the preprint. The strategy we ultimately produced (Workflow v1) is illustrated as a flowchart in the figure below. Each stage of the workflow and each new method or tool that was developed in CryoSPARC to support automated data processing, are described in the Methods section of the preprint.

Results of automated processing on GPCRs
The results demonstrated that we were able to successfully produce maps with resolution and map quality equal or better than manual processing in the repeat-target scenario, with no manual intervention. The automation strategy worked successfully even in the context of datasets containing poor quality micrographs, contaminants, aberrations, aggregation, denatured and junk particles. In multiple cases, the automated workflow yielded improved map quality and interpretability in the receptor region where a ligand is bound, demonstrating that the strategy is realistic for real-world drug discovery scenarios. Finally, the automated workflow could be completed in a day or less on a 2-GPU workstation, with further speedups possible by adding parallel compute resources.


Using the GPCR workflow file on your own GPCR datasets
The instructions below outline how to take our GPCR Workflow file and use it on your own GPCR datasets (i.e., targets of the same “class”).
Download the workflow JSON files, volumes, and masks used in the whitepaper: Automated_Workflow_materials_v1.zip
Upload these volumes to your compute setup. They do not need to be located within the project.
Import workflow JSON file into CryoSPARC using the workflow panel in the sidebar.
Select the workflow from the list, set all parameters, and choose the compute node to utilize. Parameters that might need to be adjusted (on a per dataset basis) are below:
Import movies
Movies data path
Gain reference path
Raw pixel size (Å)
Accelerating voltage (kV)
Spherical aberration (mm)
Total exposure dose (e/Å^2)
Flip gain ref and defect file in Y
Exposure group metadata (if present)
Import 3D Volumes
Path to volumes/masks for GPCR reference, junk volumes, and the receptor mask
Patch Motion Correction
Output F-crop factor
Exposure Group Utilities
Use a regular expression string
Change this regular expression string to one of the following:
If testing on an EMPIAR dataset in used in this study, refer to Table S1 for the correct regular expression to use.
If using another dataset that captures the beam image-shift group in the filename, use a regular expression that would separate them.
If beam image-shift metadata was imported with the movies, then set this job to the cluster and split method and supply the number of expected groups and other parameters that might be needed.
If beam image-shift was not used during data collection, the regular expression parameter can be left blank.
Extract from micrographs (x2)
Extraction box size
Note: if using our reference, this should be 310.6 divided by the pixel size of the motion corrected micrographs (rounded to the nearest even pixel - e.g. 361.18 —> 362).
Click on the green “Apply” button in the bottom of the workflow GUI.
Download and inspect your final volumes, and proceed with any advanced processing if desired.